11.3 Over/under temperature safety device class 2 (option)

RESET CL. 2.0

The over-/under temperature safety device (6) consists of two entry

modules located in the lateral control panel. Both modules can be

set from -50 °C / -58°F up to 200 °C / 392°F and serve to define the

maximum high and low temperature limits.

With this option, the safety controller (chap. 11.2) must be

set to maximum temperature.



(6a) Upper module: Entry of the higher limit temperature.

(6b) Lower module: Entry of the lower limit temperature.

When the temperature inside the chamber leaves this tolerance

bandwidth, the temperature control, and herewith the heating and

refrigeration, are turned off permanently.
At the corresponding entry module, the red pilot lamp K1 lights up

(K2 is without function).
The controller displays the alarm message „TEMP CLASS“. Addi-

tionally there is an audible alert, provided that the buzzer has not

been deactivated in the “User level” menu.
Let the chamber heat up or cool down to the defined safety tempera-

ture range.
The press down RESET button (5) “RESET CL 2.0” located in the

lateral control panel to re-activate the chamber. The red pilot lamp

K1 goes off.
Then reset the alarm message at the controller display by controller

button RESET (see chap. 12).

P

K2

K1

Setting limit temperatures at modules (6a) and (6b):

Press down button P

The display changes to entry mode

Enter the desired limit temperature via the arrow keys

The entered temperature value is adopted after a few

seconds. The display shows the actual temperature

again.