MK (E2.1) 10/2014

page 44/73

10. Bedew protection facility (operation line 1)

The bedew protection condensates the chamber humidity at the coldest point in order to avoid the sam-

ples becoming wet from condensation. Bedew protection is performed by the evaporator and can be pro-

grammed On/Off via operation line 1 in Manual Mode (HAND) and in Program Mode (AUTO). The other

operation lines are non-functional.

Use the bedew protection only if absolutely necessary to prevent condensation on the charging

material.

When the bedew protection is enabled (operation line 1 = On) the refrigeration machine keeps operating

within warming-up phases (On = refrigeration machine operating, Off = refrigeration machine off).

If possible, use the bedew protection only during warm-up phases. If necessary it can also be

activated during hold phases.

Do NOT use the bedew protection above a temperature set-point of +20 °C maximum.

To obtain optimal warming results without condensation on the samples, program a heating gradient of

about 0.5 °C/min.
Example:

W/°C

t/min.

00

01

02

03

-40

-20

0

20

40

60

80

120

180

380

04

ON

OFF

t/min.

Minimum

tolerance

Maximum

tolerance

Operation line 1 = bedew protection


In the program table the state of operation

line 1 for bedew protection is represented

in the column “Sk”:
On = 0000000

1

Off = 0000000

0

Example:
Operation line 1 is activated in the first

program section and remains active up to

+20 °C.
Then the refrigeration machine turns off

automatically.

PGM

08:43:55 15.12.13