Field accuracy check (acoustic calibration) – Casella CEL CEL-593 User Manual

Page 8

Advertising
background image

22. Press

until

1

1

SLM is indicated in the

left hand cursor field.

23. Press RECALL key to display the

1

1

SLM store Directory.

24. Use

to select the date and time of

the record just stored.

25. Press

to display the Header screen

for the run.

26. Press

to display the Recall data

screen 1 for the first spectrum.

27. Press

to display the Recall data

screen for the second spectrum.

6.

Field Accuracy Check (Acoustic Calibration)

28. Press ESC to return to the Status

screen.

29. Press power to switch the analyser

OFF.

You have now operated the instrument,

using all of the steps required to store and

recall data.

These are the steps required to

perform a simple noise survey.

Easy wasn’t it ?

Setting the clock and performing a

calibration can be carried out in any

Application and will be valid for all

Applications.

With batteries, preamplifier,

microphone and interface installed

according to Chapter 3 above, proceed as

follows.

1. Press the power key to switch the

instrument ON.

2. Wait for the self test screens to clear

and show the Configuration screen.

3. When the date and time are correct, go

to step 8.

4. When the date and time are incorrect

use the SET CLOCK option to display

the Set Clock screen.

5. Set the date and time using cursor and

value keys.

6. Press the SET key.

7. Press ESC key to return to the

Configuration screen.

8. Use the CAL option to display the first

Calibrate screen.

The Calibration screen can also be

reached via the CAL option on the

Options screen, see Chapter 10.

9. When the previous calibration is

acceptable, use the ABORT option and

go directly to step 14.

When the previous calibration is not

acceptable, use the CAL option to

produce the second Calibrate screen.

10. Fit a CEL-284/2 Calibrator over a

Type 1 microphone and either a

CEL-282 or a CEL-284/2 Calibrator over

a Type 2 microphone, making sure the

microphone is correctly seated against

the shoulder in the cavity (Figure 3).

11. Switch the calibrator ON.

These calibrators generate a nominal level

of 114.0 dB at 1 kHz.

12. For the CEL-192, CEL-250 and MK 250,

if necessary use the LEVEL keys to

adjust the sensitivity until 114.0 dB is

indicated.

For the MK 301, adjust the indicated

level to 94.0 dB.