Casella CEL Software for the Tuff sampling pumps User Manual

Page 11

Advertising
background image

I. SET REQUIRED FLOW RATE

(Fig. 12)


Note: “Set Flow” and “Reset” options are not available in Partial Lock mode.
From START, press UP or DOWN arrow to go to SET FLOW. Press ENTER to access
Set Flow options. Use UP/DOWN keys to adjust the flow setting in 0.5 l/min
increments. (Please note Tuff 3 will only adjust to maximum of 3l/min whilst Tuff 4
models will adjust up to 4.5L/min flow rates.)
Press ENTER to continue.
Pump should then be connected to external flow calibration device.
For External Calibration details refer to

Chapter 6

.


Use UP/DOWN keys to calibrate the pump flow setting

.

Figure 12 Calibration Screens

J. POWERING OFF

(Fig. 13)

To stop the pump at any time, press and hold press the ON/OFF Key. A “countdown”
will be displayed and the pump will switch off.

Figure 13 Power OFF Screens

Casella CEL Limited

11

Advertising
This manual is related to the following products: