DAQ signal timing

The DAQ timing signals are:

 SCANCLK
 A/D START TRIGGER
 A/D STOP TRIGGER
 STARTSCAN
 SSH
 A/D CONVERT
 A/D PACER GATE
 A/D EXTERNAL TIMEBASE
 A/D STOP

SCANCLK signal

SCANCLK is an output signal that may be used for switching external multiplexers. It is a 400 ns wide pulse
that follows the CONVERT signal after a 50 ns delay. This is adequate time for the analog input signal to be
acquired so that the next signal may be switched in. The polarity of the SCANCLK signal is programmable. The
default output pin for the SCANCLK signal is AUXOUT2, but any of the AUXOUT pins may be programmed
as a SCANCLK output.

CONVERT

SCANCLK

t

d

t

d

t

w

t

d

= 50 ns

t

w

= 400 ns

Figure 4. SCANCLK signal timing

A/D START TRIGGER signal

Use the A/D START TRIGGER signal for conventional triggering (when you only need to acquire data after a
trigger event). Figure 5 shows the A/D START TRIGGER signal timing for a conventionally triggered
acquisition.

A/D Start Trigger

Start Scan

Convert

1

2

3

4

0

Scan Counter

Figure 5. A/D START TRIGGER

— Data acquisition example for conventional triggering

The A/D START TRIGGER source is programmable and may be set to any of the AUXIN inputs. The polarity
of this signal is also programmable to trigger acquisitions on either the positive or negative edge.

The A/D START TRIGGER signal is also available as an output and can be programmed to appear at any of the
AUXOUT outputs. See Figure 6 and Figure 7 for A/D START TRIGGER input and output timing
requirements.