Factory-set Test Sequences

94

HH/PORT 1

110V
HH/PORT 2

SKIP 2.0M

Ω SKIP 0.25mA SKIP

110V
MOV/STAT 1

0.1

Ω 1.0MΩ 3.5mA SKIP SKIP

110V
MOV/STAT 2

SKIP 2.0M

Ω SKIP 0.25mA SKIP

110V LEAD
(3M)

0.08

Ω 1MΩ SKIP SKIP SKIP

110V LEAD
(6M)

0.2

Ω 1MΩ SKIP SKIP SKIP

110V LEAD
(12M)

0.3 1M

Ω SKIP SKIP SKIP

All tests have visual test included

Extension lead tests include a Polarity check

Factory-set Test Sequences

110V
HH/PORT 2

SKIP 2.0M

Ω SKIP 0.25mA SKIP

110V
MOV/STAT 1

0.1

Ω 1.0MΩ 3.5mA SKIP SKIP

110V
MOV/STAT 2

SKIP 2.0M

Ω SKIP 0.25mA SKIP

110V LEAD
(3M)

0.08

Ω 1MΩ SKIP SKIP SKIP

110V LEAD
(6M)

0.2

Ω 1MΩ SKIP SKIP SKIP

110V LEAD
(12M)

0.3 1M

Ω SKIP SKIP SKIP

All tests have visual test included

Extension lead tests include a Polarity check