Model nga2000 hfid – Emerson Process Management ROSEMOUNT NGA2000 HFID User Manual

Page 32

Advertising
background image

Instruction Manual

748297-E
June 2003

3-2 Startup and Operation Rosemount Analytical Inc. A Division of Emerson Process Management

Model NGA2000 HFID

EPROM test - Checks the

EPROM on the Analysis Com-
puter PCB.

RAM test - Checks the RAM on

the Analysis Computer PCB.

Power supply test - Verifies that

all internal DC voltages are within
the required tolerances.

Network test - Checks the inter-

nal network interface.

20 bit ADC test - Checks the

20-bit ADC on the Analysis Com-
puter PCB by sending a DC sig-
nal through the Preamp PCB and
reading the signal back with the
20-bit ADC.

12 bit ADC test - Checks the

12-bit ADC on the Analysis Com-
puter PCB by sending a DC sig-
nal and reading the signal back
with the 12-bit ADC.

Power Supply PCB test -

Checks the presence of the
Power Supply PCB by activating
the 3-way air solenoid.

Safety PCB test - Checks the

presence of the Safety PCB by
sending a command and reading
it back.

Case temperature test - Com-

pares the temperature read be-
tween the Preamp temperature
sensor and the case temperature

sensor. They must be within 10

°C

of each other. This test some-
times fails if the case is opened.
The sensor in the Preamp will
take longer to cool off since it is in
an enclosure. Re-running the
self-test after thermal equilibrium
will produce a positive result if the
sensors are working properly.

Oven/Sample Temperature test

- Compares the temperature read
between the sample temperature
sensor and the oven temperature
sensor. They must be within 50

°C

of each other.

• The self-test can be repeated at

any time by activating the TEST
softkey in the Self Test Results
menu.

8. Set the desired oven setpoint in the

range of 93

°C to 204°C (200°F to

400

°F).

9. Wait for the Purge Air green LED to il-

luminate.

10. Introduce the remaining supply gases.

Perform leak check. (See Section 1-5
Specifications on page 1-4.)

11. Set and verify the internal gas pres-

sures.

INTERNAL PRESSURE REGULATOR

TYPICAL OPERATING PRESSURES

Burner Air

965 to 1103 hPa-gauge (14 to 16 psig)

Fuel

1516 to 1723 hPa-gauge (22 to 25 psig)

Sample (non-adjustable)

206 to 290 hPa-gauge (3.0 to 4.0 psig)

Purge air of the following specifications must be present:

Flow:

16 to 18 L/min.

Supply Pressure: 689 to 1378 hPa-gauge (10 to 20 psig)