Fuel in 2-way solenoid valve, Burner air in filter, Regulated air in check valve – Emerson Process Management ROSEMOUNT NGA2000 HFID User Manual

Page 66: Model nga2000 hfid

Advertising
background image

Instruction Manual

748297-E
June 2003

4-28 Maintenance and Troubleshooting

Rosemount Analytical Inc. A Division of Emerson Process Management

Model NGA2000 HFID

Fuel In 2-Way Solenoid Valve

1. Disconnect wiring solenoid valve wir-

ing connector, note location.

2. Inside the analyzer module, discon-

nect the tube going to the connector
on the “out” port of the solenoid valve.

3. On the rear of the analyzer module at

the fuel in port:

a.

Disconnect the fuel in tube.

b.

Remove nuts and washers.

4. Remove solenoid valve from analyzer

module

5. Remove the fittings from the solenoid

valve and replace the Teflon pipe
thread tape.

6. Verify that body of replacement sole-

noid valve is oriented as shown in
Figure 4-22 on page 4-27. If not, ro-
tate till wires are in-line with “out” port.

7. Install fittings into replacement sole-

noid valve, re-install in analyzer mod-
ule.

Burner Air In Filter

1. Leaving the bulkhead fitting secured

to the rear panel, remove the tubes,
nuts and ferrules from the fitting.

2. Insert a clean, rigid piece of tube or

rod (smaller than .25 inch diameter)
into the bulkhead fitting to force out
the filter disc.

3. Install the replacement filter in the

same manner, through the rear of the
bulkhead fitting.

4. Re-connect

tubes.

Heated Bypass Sample Out and Heated
Sample In Restrictors

1. On the outside of the rear panel, dis-

connect tube and remove nut.

2. Insert a small spade screwdriver into

the bulkhead and remove the restric-
tor.

3. Install in reverse order.

Regulated Air In Check Valve

1. Disconnect tube at elbow.

2. Remove check valve from female

connector.

3. Remove elbow from check valve.

4. Add Teflon pipe thread tape to check

valve threads.

5. Install elbow onto check valve.

6. Install check valve into female con-

nector, verifying orientation of elbow
fitting as shown in Figure 4-22 on
page 4-27.

Advertising