4-8 Maintenance and Troubleshooting

Rosemount Analytical Inc. A Division of Emerson Process Management

Model NGA2000 HFID

Replacing Burner Jets

Disassemble the burner only if contami-
nants are evident. Combustion products
or other contaminates which accumulate
inside the burner may form electrical
leakage paths between the collector and
the burner contact, resulting in noisy read-
ings.

If the analyzer module is to be operated at
the highest sensitivity, traces of such con-
taminates can cause erroneous readings.
For best performance, replace the burner
jet follows:

WARNING

BURNER CONTAMINATION

Do not handle internal parts of the
burner with bare hands. All tools used
for maintenance must be free of con-
taminates.

1. Remove oven from analyzer module

per Section 4-2a on page 4-3.

2. Remove burner/thermal block from

oven per 4-2b on page 4-4.

3. Remove thermal block from burner

per Section 4-2c on page 4-7.

4. Refer to Figure 4-7A below. Remove

screws (2) holding burner cap re-
tainer, remove retainer.


























Figure 4-7.

Burner Disassembly

Burner Cap

Burner Cap
Retainer

Gasket
656931

Burner Base

Burner Base

Combustion
Chamber

A

B

O-Ring
904373