Maintenance and Troubleshooting 4-7

Model NGA2000 HFID

c.

Burner Internal Components

WARNING

BURNER CONTAMINATION

Do not handle internal parts of the
burner with bare hands. All tools used
for maintenance must be free of con-
taminates.

Disassembly of Burner/Thermal Block

1. Remove oven from analyzer module

per Section 4-2a on page 4-3.

2. Remove burner/thermal block from

oven per Section 4-2b on page 4-4.

3. Refer to Figure 4-6 below. Discon-

nect sample capillary nut at base of
burner.

4. Remove screw securing thermal block

to burner.

Carefully pull burner away from thermal
block.

























Figure 4-6.

Burner/Thermal Block Disassembly











Sample capillary

Sample capillary
input

Burner

Thermal Block