Maintenance and Troubleshooting 4-11

Model NGA2000 HFID

Sample RTD
657061

Cartridge Heater
659643

Thermostat
657065

d.

Thermal Block

The sample RTD can be replaced with the
thermal block attached to burner and
mounted in oven. The cartridge heater
and thermostat are also replaceable with
thermal block secured to burner, but must
be removed from the oven.

Sample RTD

1. Refer to Figure 4-4A on page 2-1.

Remove the four screws securing the
oven cover, remove cover.

2. Disconnect the sample RTD wiring

connector, note location.

3. Refer to Figure 4-9 below. Remove

the two screws securing the sample
RTD, pull sample RTD out.

4. Install replacement sample RTD, se-

cure with screws.

5. Attach sample RTD wiring connector.

6. Re-attach oven cover.

























Figure 4-9.

Thermal Block – Sample RTD, Cartridge Heater and Thermostat


Cartridge Heater

1. Remove oven from analyzer module

per Section 4-2a on page 4-3.

2. Remove burner/thermal block from

oven per Section 4-2b on page 4-4.

3. Refer to Figure 4-9 above. Loosen

retaining set screw, pull out cartridge
heater.

4. Install replacement cartridge heater,

snug down set screw.

5. Install burner/thermal block into oven.

6. Install oven into analyzer module.