Preamp assembly, 15 preamp assembly installation, Model nga2000 hfid – Emerson Process Management ROSEMOUNT NGA2000 HFID User Manual

Page 55

Advertising
background image

Instruction Manual

748297-E

June 2003

Rosemount Analytical Inc. A Division of Emerson Process Management

Maintenance and Troubleshooting 4-17

Model NGA2000 HFID

Preamp Assembly

1. Per Section 4-2e on page 4-13, re-

move securing hardware from elec-
tronics assembly and lay on bench.

2. Disconnect and note location of ca-

bles.

3. Remove the two screws and washers

from the top bracket and slide the
preamp assembly out.

4. Remove the lower bracket from the

preamp assembly and install on re-
placement preamp assembly.

5. Slide replacement preamp assembly

into top bracket and secure with
mounting hardware.

6. Re-connect

cables.


































Figure 4-15.

Preamp Assembly Installation

Bottom Bracket

Top Bracket

Preamp Assembly
656945

Advertising