E. electronics assembly, 11 removing electronics assembly from chassis, Model nga2000 hfid – Emerson Process Management ROSEMOUNT NGA2000 HFID User Manual

Page 51

Advertising
background image

Instruction Manual

748297-E

June 2003

Rosemount Analytical Inc. A Division of Emerson Process Management

Maintenance and Troubleshooting 4-13

Model NGA2000 HFID

Electronics Assembly
656943

Hex Nut

Screw

e. Electronics

Assembly

The electronics assembly must be re-
moved from the chassis if replacement of
any of the following components is nec-
essary:

Power Supply Board

Safety Board

Computer Analysis Board

Preamp Assembly

Sensor Board

Case Temperature Sensor

Case Pressure Switch

1. Remove the hex nut and screw as

shown in Figure 4-11 below.

2. Lay electronics assembly on bench,

do not disconnect cables or tubing.

































Figure 4-11.

Removing Electronics Assembly from Chassis

Advertising