C. drift, D. noise, Model nga2000 hfid – Emerson Process Management ROSEMOUNT NGA2000 HFID User Manual

Page 68

Advertising
background image

Instruction Manual

748297-E
June 2003

4-30 Maintenance and Troubleshooting

Rosemount Analytical Inc. A Division of Emerson Process Management

Model NGA2000 HFID

c. Drift

1. Verify that the sample, burner air,

and fuel supply pressures are
constant.

2. Check that the tubing, regulators,

pumps, fittings, and valves are
clean of hydrocarbons.

3. Verify that the oxygen level in the

burner air and sample are con-
stant.

4. Verify the THC level is correct for

the burner air and fuel supply.

5. Check that the ambient tempera-

ture is changing <10

°C per hour.

6. Verify the burner is clean.

7. Verify temperature of the sample

gas, case, burner, and oven has
stabilized.

8. Verify the Preamp PCB is clean.

9. Verify atmospheric pressure at

burner exhaust is constant.

10. Verify purge gas pressure is con-

stant.

11. Verify burner has been on and

stabilized.

12. Check for gas leaks.

d. Noise

1. Check that the burner exhaust is

free from water condensation.

2. Verify connection to the collector is

correct.

3. Verify connection to the polarizing

voltage is correct.

4. Check the ambient temperature is

changing <10

°C per hour.

5. Verify the +24VDC is clean and

grounded properly.

6. Verify there are no strong magnetic

fields near.

7. Check for excessive vibration.

8. Verify burner exhaust is vented to a

constant atmospheric pressure.

9. Verify bypass line is vented to a

constant atmospheric pressure.

10. Verify purge out port vented to a

constant atmospheric pressure.

11. Verify the collector wires are routed

away from the heater.

12. Verify the collector wires are clean

and not damaged.