4-10 Maintenance and Troubleshooting

Rosemount Analytical Inc. A Division of Emerson Process Management

Model NGA2000 HFID

Installation

WARNING

BURNER CONTAMINATION

Do not handle internal parts of the
burner with bare hands. All tools used
for maintenance must be free of con-
taminates.

1. Install

new lower gasket, jet assembly

and upper gasket into burner base,
finger-tight jet nut.

2. Install new sample jet (with gasket)

and tighten.

3. Tighten jet nut.

4. Install air baffle per Figure 4-8 on

page 4-9.

Note

Incorrect installation of air baffle will
cause ignition failure.

5. See Figure 4-7B on page 4-8. Insert

new o-ring into burner base.

6. Set combustion chamber into burner

base being careful not to move air
baffle
.

7. See Figure 4-7A on page 4-8. Insert

new gasket on combustion chamber,
install burner cap and burner cap re-
tainer, torque screws to 6 inch lbs.