Operation Procedure

(Without data management software CS-S10w)

Setting Necessary for Object Color Measurement

1.

Set one or more tungsten lumps

or equivalent as illumination

source toward white calibration

plate as in the right illustration.

• Set this instrument vertical
to white calibration plate.
• Keep the angle between
illumination light source and
white calibration plate at 45˚.

White Calibration

2.

Perform user calibration.

See p.64 for details.

Measurement of Object

3.

Set object on the same position and at the
same angle as white calibration plate.

4.

Now, measurement can be done.

Illumination

Light Source

Illumination

Light Source

White calibration plate