Internal Sync Measurement Mode Setting

Internal sync measurement mode refers to measurement mode where measurement is
made in the same timing as periodical light source pulse frequency, such as vertical syn-
chronization frequency for display.

∗ Setting at the shipment from factory : NO SYNC

Operation Procedure

1.

Press

ESC

key when menu or target

value setting menu appears.

Measurement screen appears on LCD screen.

2.

Make sure that [ ] is shown on bottom

left of measurement screen indicating

it has not been set for SHIFT mode.

If [SFT] appears instead, it means that
instrument is to operate in SHIFT mode.
Press

SHIFT key to cancel SHIFT mode.

3.

Press

MENU

key.

Menu 1/4 screen appears on LCD screen.

4.

Press either

or

key to select

[SYNC MODE] and then

ENTER

key.

< SYNC MODE> screen appears on LCD screen.