Konica Minolta Chroma Meter CS-200 User Manual

Page 116

Advertising
background image

114

De

scr

ipti

on

Error

No.

Symptom

Item to Check

Corrective Action

Page to

Refer

6

Measure-
ment value
appears as
“------“.

Does data exist?

This appears when there is no data
in measurement value, stored data,
calibration value and target color.

-

Doesn’t color space
become color tempera-
ture?

This appears when color tempera-
ture cannot be converted for display.
Available display range is as follows:
  

2300 T 20000(K)  |∆uv|<0.1

34

Doesn’t measurement
value become larger by
user calibration?

This appears when luminance value
exceeds the available display area in
this instrument.

-

Do you set target color
for color difference
measurement?

This appears in color difference dis-
play when target color is not set.

78

7

Measurement
values vary.

Is object to measure
stable?

Keep object to measure stable.

-

Isn't object to measure
of low luminance?

Repeatability of x, y worsens if ob-
ject of low luminance is measured.
It becomes worse especially when
measurement angle is 0.2˚ or 0.1˚.
Also it becomes worse especially
when measurement time is FAST
or Super-FAST. Measure in slower
measurement time.

4

28

Is measurement sync
frequency appropriate
when measuring dis-
play?

Set appropriate measurement sync
frequency for proper measurement.

30

Haven't ambient tem-
perature and humidity
rapidly changed?

Conduct measurement under envi-
ronment free from such fluctuation.

3

8

Strange

measurement

value ap-

pears.

Is objective lens clean? Wipe off with dry and soft cloth or lens

cleaning paper.

106

Has user calibration

been performed cor-

rectly?

Redo user calibration.

64

Haven't you put close-

up lens on?

Select lens type setting correspond-
ing to applied close-up lens.

62

Haven't you put ND

filter on?

Select lens type setting correspond-
ing to applied ND filter.

40