Error message

Cause (Description)

Corrective Action

7

OVER

Luminance of measuring
object is higher than avail-
able measurement range.

• Use ND filter and redo measure-

ment.

• If symptom does not improve,

please contact the nearest KONI-
CA MINOLTA SENSING authorized
service facility.

8

OFFSET ERROR

Zero calibration has not
been correctly performed.

• Reset the power and redo meas-

urement.

• If symptom does not improve,

please contact the nearest KONI-
CA MINOLTA SENSING authorized
service facility.

9

EXCESSIVE LUMI-
NANCE VARIATION

Luminance change for
measuring object is large.

• Place this instrument on stable

surface and redo measurement.

• If symptom does not improve,

please contact the nearest KONI-
CA MINOLTA SENSING authorized
service facility.

10

VIEWING-ANGLE
SELECTOR ERROR

Measurement was per-
formed when measurement
angle selector was in wrong
position, or measurement
angle selector was changed
during measurement.

• Switch measurement angle selec-

tor and check that measurement
angle appears in screen (------ is
not displayed), and then redo
measurement. In addition, do not
operate measurement angle selec-
tor during measurement.

• If symptom does not improve,

please contact the nearest KONI-
CA MINOLTA SENSING authorized
service facility.

11

ROM ERROR

Data stored in ROM is
broken.

• Do not switch OFF (O side) while

storing data or changing setting, or
when message “PLEASE WAIT...”
appears.

• If symptom does not improve,

please contact the nearest KONI-
CA MINOLTA SENSING authorized
service facility.