Error messages – Konica Minolta Chroma Meter CS-200 User Manual

Page 112

Advertising
background image

110

De

scr

ipti

on

Error Messages

Error message appears on LCD screen for incorrect operation of this instrument through
key. Below table shows type of error message, its description and corrective action
respectively.

Error message

Cause (Description)

Corrective Action

1

BATTERY OUT

Battery voltage decreases.

• Switch this instrument OFF (O

side), and replace with new batter-
ies or use AC adapter.

2

DATA SET ERROR Entered numerical value is

out of range.

• Enter again. Note that measure-

ment value should completely
satisfy the ranges below:

0<x<1,

0<y<1,

0<x+y 1,

0<X 99999000000,

0<Y 99999000000,

0 Z 99999000000

3

CH00 DATA
NOT SETTA-
BLE BY USER

When CH00 is selected,
calibration channel is go-
ing to be operated or target
color is going to be set.

• Select channel except CH00 and

redo the operation.

4

INCORRECT
OBSERVER
CONDITION

Measurement data measured
with different observer angle
from that for channel is going
to be used for user calibration
for that channel, or for target
color setting.

• Select measurement data meas-

ured with channel and observer
angle or reset calibration channel,
and redo setting.

5

NO DATA

Without measurement data,
user calibration is going
to be performed, or target
color to be set.

• Select data to measure or meas-

urement data, and redo setting.

6

UNDER

Luminance of measuring
object is lower than avail-
able measurement range.

• Remove lens cap and check if

luminance of measuring object
is within available measurement
range, and then redo measure-
ment.

• Check if user calibration is cor-

rectly performed, and redo meas-
urement.

• Check if settings for measurement

angle, observer angle and lens are
correct, and redo measurement.

• If symptom does not improve,

please contact the nearest KONI-
CA MINOLTA SENSING authorized
service facility.