Selecting Measurement Time

Select measurement time depending on purpose.
7 modes are available for measurement time.
Select the mode with long measurement time when repeated accuracy is required such
as when measuring object of low luminance.

∗ Setting at the shipment from factory : AUTO.

Zero Calibration

Measurement Time calculated by formula
(Time for integration x 2 +Time to open/close

shutter + Time for calculation) indicating time

needed for actual measurement.

Time for integration
Time for sensor to measure light
indicating “exposure time”.

Super-FAST Not be done every time ∗1

0.5s approx.

300 ms

FAST

Done every time
automatically

1s approx.

300 ms

SLOW

3s approx.

1300 ms

Super-SLOW

12s approx.

1300 ms x 4 times

LTD.AUTO 2

1s or 3s approx.

300 ms or 1300 ms

AUTO 3

1s to 60 s

MANUAL

1s to 60 s (Every 1s)

∗ Zero calibration is performed automatically one minute after previous measurement.
∗ 2 Selects FAST/SLOW automatically depending on luminance. Luminance for which FAST/SLOW is switch-

able is as follows for measuring angle 1˚.
FAST → SLOW: L

v

=20 cd/m

2

or lower (Measurement is redone in SLOW mode.)

SLOW → FAST: L

v

=40 cd/m

2

pt or higher

∗ 3 Selects the measurement time automatically depending on luminance.

For measuring angle 1°, when luminance is higher than 20 cd/m

2

, the measurement time switches as LTD.

AUTO mode. When luminance is 20 cd/m

2

or lower, the measurement time is calculated depend on luminance

as following examples. The higher limit is 60 s.

L

V

20.0 cd/m

2

15.0 cd/m

2

10.0 cd/m

2

6.6 cd/m

2

4.6 cd/m

2

Measurement Time

3 s approx.

7 s approx.

14 s approx.

30 s approx.

60 s approx.

Operation Procedure

1.

Press

ESC

key when menu or target

value setting menu appears.

Measurement screen appears on LCD screen.

2.

Make sure that [ ] appears on bottom

left of measurement screen indicating

it has not been set for SHIFT mode.

If [SFT] appears instead, it means that
instrument is to operate in SHIFT mode.
Press

SHIFT key to cancel SHIFT mode.

2

3

1

7

5

4, 6

2

3

1

7

5

4, 6

M

E

A

S

S

P

E

E

D

CH00:

<

DEFAULT

>

Lv ------

cd/m

2

x ------

y ------

<

MODE

>

SINGLE

[ ]  [ 2° ]

<

ANGL

>

1° VIEW

<

SPD

>

AUTO

<

SYNC

>

NO SYNC

<

LENS

>

STANDARD

<

MEM

>

AUTO NUM

<

MEMORY DATA

>

M

OOO

:(NO DATA)

Lv ------

x ------

y ------

Advertising