Lincoln Impinger Conveyor Ovens 1100 Series User Manual

Page 29

Advertising
background image

Flame will not stay on

Flame Sensor

To check for flame sensor operation,

connect a digital multimeter (capable

of measuring D.C. micro amps)

between the flame sensor wire and

terminal “F” on the Honeywell Ignition

Control or terminal 1 on Landis & Gyr

Ignition Control. Sensor current from

Honeywell is 2 to 4 D.C. micro amps.

Sensor current from Landis & Gyr is

3 micro amps minimum.

If these readings are not achieved

replace ignitor/sensor assembly.

Also check for any type of damage to

flame sensor wire and connections.

NOTE: The D.C. micro amp test

must be conducted with the oven in

low flame (bypass) operation.

Power Supply

Turn the temperature control to

the lowest setting. If there is

sufficient micro-amp current, but

the flame will not stay lit, check

for proper polarity of the power supply

Ignition Control

If there is sufficient flame sensor

current, but the burner will not

remain ignited, check the reset button

on ignition control. If all above are

okay, replace ignition control.

Pilot lamp is on, but

Temperature Control

Check for 120 VAC across terminals

no main flame

10 and 11 on temperature control

board. If no voltage is present, check

wiring back to the Main Fan Switch.

Turn the temperature adjustment

knob to the maximum temperature

position and check for 120 VAC at

the load terminal (number 9), and

neutral (number 10).

If 120 VAC is present and unit is not

heating, refer to “Solenoid Valve” for

next check. If 120 VAC is not present,

proceed with next step.

Thermocouple Probe

With power on and thermocouple

leads attached to board, measure

the millivolt output of these leads.

Refer to the thermocouple chart in

Section D for proper readings.

Temperature Control

WITH POWER OFF: Remove the

Potentiometer

potentiometer leads from the temperature

control board and measure ohms from black

to whiteleads. The reading should be from

approximately 0 to 2.5K ohms as the dial of the

potentiometer is turned through its full rotation.

From black to red leads, the reading should be

approx.2.5K ohms and remain steady throughout

the full rotation of the potentiometer.

If the thermocouple probe and the

control potentiometer check good,

Impinger II - 1100 Series Service Manual - International

29

Advertising