Resetting calibration data – Perle Systems IOLAN SCS User Manual

Page 314

Advertising
background image

Calibrating I/O

314

IOLAN SDS/SCS/STS User’s Guide, Version 3.6

You are now prompted to apply 20 mA to the positive (+) and negative (-) terminals. Once that is
done, click

Yes

to proceed.

Once calibration is successfully completed, click

OK

to finish the process.

Resetting Calibration Data

You can reset the I/O channels calibrations to the factory calibrations in DeviceManager by selecting

Tools

,

I/O Channels

,

Reset Calibrate Data

or in WebManager by selecting

Administration

,

Reset

,

I/O Calibration

.