Chapter 4 Signal Connections

PC-LPM-16/PnP User Manual

4-6

National Instruments Corporation

Figure 4-2.

Analog Input Signal Connections

Digital I/O Signal Connections

See Table 4-1 for the digital I/O pin descriptions.

The following specifications and ratings apply to the digital I/O lines:

Absolute maximum

+7.0 V with respect to DGND

voltage input rating

-0.5 V with respect to DGND

Digital input compatibility

TTL-compatible

Input current (high or low level)

±

10

µ

A

Digital output compatibility

TTL-compatible

Output current source capability 8 mA, at V

OH

= 2.7 V

Output current sink capability

6 mA, at V

OL

= 0.5 V

+

+

-

-

V

M

ACH<0..15>

Input Multiplexer

Operational

Amplifier

I/O Connector

3

4

5

18

1, 2

PC-LPM-16PnP

AIGND

-

+

-

Signal

Source

Measured

Voltage

Vs

1

+

-

V

S2

+

V

S3