Command register 2 – National Instruments PC-LPM-16/PnP User Manual

Page 79

Advertising
background image

Appendix D Register-Level Programming

National Instruments Corporation

D-7

PC-LPM-16/PnP User Manual

Command Register 2

Command Register 2 contains only one bit that enables the auto-calibration operation of
the ADC.

Address:

Base address + 07 (hex)

Type:

Read-and-write

Word Size:

8-bit

Bit Map:

Bit

Name

Description

7–3

0

Reserved bits—These bits must be set to zero for
future board compatibility.

2

SCANORDER

Scan Order Bit—If this bit is cleared, the scan order is
from the channel in Command Register 1 MA<3..0>
to channel 0. The power-on value is 0. If this bit is set,
the scan order begins with channel 0 and ends with the
channel number in MA<3..0>. This bit is cleared upon
power up. To ensure proper scanning, this bit should
be correctly programmed before writing to the
SCANEN* and Channel Selection bits in Command
Register 1. This bit is only present on the
PC-LPM-16PnP.

Note:

The UP function is not yet supported by NI-DAQ. NI-DAQ will support the
UP function in a future release.

1

DISABDAQ

Disable Data Acquisition Bit—This bit is used to
disable the data acquisition operation. The power-on
value is 0. Upon startup, this bit is cleared and, as a
result, the data acquisition operation is enabled.
Writing a one to this bit disables both A/D conversion
source signals OUT0* and EXTCONV*.

7

6

5

4

3

2

1

0

0

0

0

0

0

SCANORDER

DISABDAQ

CALEN

a.Book : l.Appendix D Page 7 Wednesday, November 20, 1996 6:36 PM