Chapter 2

Connecting Signals

© National Instruments Corporation

2-11

NI PXI-4224 User Manual

Figure 2-9. Shielded Grounded Signal Source Connection Using a Terminal Block

Floating Signal Source Connection

Figures 2-2, 2-4, 2-6, and 2-8 illustrate floating signal source connections.
In this configuration, the signal source being measured is a floating signal
source, such as a battery. A floating signal source is not connected in
any way to the building ground system.

To connect a floating signal source connection to the NI PXI-4224, the
signal (V

SIG

+) is connected to the NI PXI-4224 channel (AI X +). The

signal reference (V

SIG

–) is connected to the channel reference (AI X –).

Twisted-Pair

Wiring

AI 0 +

AI 0 –

CH 0

Terminal Block

AI 7 +

AI 7 –

CH 7

Shielding

+

V

SIG

V

SIG

Ground

Reference

Signal Source