Appendix B

Timing Signal Information

NI PXI-4224 User Manual

B-10

ni.com

AI HOLD COMPLETE Signal

AI HOLD COMPLETE is an output-only signal that generates a pulse with
the leading edge occurring approximately 50 to 100 ns after an A/D
conversion begins. The polarity of this output is software configurable, but
the polarity is typically configured so that a low-to-high leading edge can
clock external analog input multiplexers that indicate when the input signal
has been sampled and can be removed. This signal has a 400 to 500 ns pulse
width and is software enabled. Figure B-12

shows the timing for

AI HOLD COMPLETE.

Note

The polarity of AI HOLD COMPLETE is not software selectable when

programmed using NI-DAQmx. It is a positive polarity pulse.

Figure B-12. AI HOLD COMPLETE Signal Timing

t

w

= 400 to 500 ns

t

d

= 50 to 100 ns

AI CONV CLK

AI HOLD COMPLETE

t

d

t

w