Figure B-10. AI CONV CLK Output Signal Timing
The ADC switches to hold mode within 60 ns of the selected edge. This 
hold-mode delay time is a function of temperature and does not vary from 
one conversion to the next. Separate the AI CONV CLK pulses by at least 
one conversion period.
The NI PXI-4224 sample interval counter generates AI CONV CLK 
unless you select an external source. The AI SAMP CLK signal starts the 
counter, which counts down and reloads itself until the scan finishes. The 
counter then reloads itself in preparation for the next AI SAMP CLK pulse.
A/D conversions generated by an internal or external AI CONV CLK 
signal are inhibited unless they occur within a sequence. Scans occurring 
within a sequence can be gated by either the hardware AI PAUSE TRIG 
signal or the software command register gate.
AI PAUSE TRIG Signal
PFI 0, PXI_Trig<0..5>, or PXI_Star can externally input the 
AI PAUSE TRIG signal, which is not available as an output on the 
I/O connector. AI PAUSE TRIG can mask off scans in a sequence. 
You can configure the pin you select as the source for AI PAUSE TRIG in 
level-detection mode. You can configure the polarity selection for the pin 
as either active high or active low. 
In level-detection mode, the AI SAMP CLK signal is masked off and no 
scans can occur.
AI PAUSE TRIG can neither stop a scan in progress nor continue a 
previously gated-off scan. In other words, once a scan has started, 
AI PAUSE TRIG does not gate off conversions until the beginning of the 
next scan. Conversely, if conversions are gated off, AI PAUSE TRIG does 
not gate them back on until the beginning of the next scan.
t
w
= 50 to 100 ns
t
w