External calibration, External calibration -13 – National Instruments Isolated Analog Input Device NI PXI-4224 User Manual
Page 49
 
Chapter 5
Using the NI PXI-4224
© National Instruments Corporation
5-13
4.
Right-click the NI PXI-4224 and select Self-Calibrate.
5.
A dialog box opens indicating that the NI PXI-4224 is self-calibrating.
6.
When the dialog box closes, the NI PXI-4224 is successfully 
self-calibrated.
Note
The NI PXI-4224 also can be self-calibrated programmatically by using
DAQmx
Self Calibrate.vi
in LabVIEW.
The results of an internal calibration are stored in the NI PXI-4224 flash 
memory so that the CalDACs are automatically loaded with the newly 
calculated calibration constants the next time the NI PXI-4224 is 
powered on.
Performing a self-calibration at the operating temperature of your 
application will ensure the NI PXI-4224 meets the specifications in 
Appendix A, 
External Calibration
You can download all available external calibration documents by going to
ni.com/calibration
and clicking Manual Calibration Procedures.
NI recommends you perform an external calibration once a year.