Figure 4-3. ai conv clk signal routing, Programmable function inputs, Programmable function inputs -6 – National Instruments Isolated Analog Input Device NI PXI-4224 User Manual

Page 33: Figure 4-3, Ai conv clk signal routing -6

Advertising
background image

Chapter 4

Theory of Operation

NI PXI-4224 User Manual

4-6

ni.com

Figure 4-3. AI CONV CLK Signal Routing

Figure 4-3 shows that AI CONV CLK can be generated from a number of
sources, such as the external signals PFI 0, PXI_Trig<0..5>, and PXI_Star,
and the Ctr 0 Internal Output.

Programmable Function Inputs

PFI 0 is connected to the front SMB connector of the NI PXI-4224.
Software can select PFI 0 as the external source for a given timing signal.
Any timing signal can use the PFI 0 pin as an input, and multiple timing
signals can simultaneously use the same PFI. This flexible routing scheme
reduces the need to change physical connections to the I/O connector for
different applications. Refer to Table 4-2 for information regarding the
available PFI 0 signals.

PXI Trigger<0..5>

PFI 0

AI CONV CLK

Ctr 0 Internal Output

PXI Star

Advertising