Ground-referenced signal connection, Recommended) -12 – National Instruments Isolated Analog Input Device NI PXI-4224 User Manual
Page 24
 
Chapter 2
Connecting Signals
2-12
ni.com
Ground-Referenced Signal Connection
Figures 2-3, 2-5, 2-7, and 2-9 illustrate the ground-referenced signal 
connection. In this configuration, the voltage source being measured is 
referenced to its own ground reference that is connected through a 
conductive path to the instrument ground reference. For example, the path 
can be through a common earth ground or through the power line ground.
To connect a ground-reference signal source to the NI PXI-4224, the signal 
(V
SIG
+) is connected to the NI PXI-4224 channel (AI X +). The signal
reference (V
SIG
–) is connected to the channel reference (AI X –).
Shielded Ground-Referenced Signal Connection 
(Recommended)
Figures 2-5 and 2-9 illustrate shielded ground-referenced signal 
connections. The connection to this signal source is identical to the 
ground-referenced signal connection with the addition of shielding 
around the field wiring. The shielding is grounded at the signal source 
ground (V
SIG
Ground Reference). Connect the signal (V
SIG
+) to the
NI PXI-4224 channel (AI X +). Connect the signal reference (V
SIG
–) to the
channel reference (AI X –).
This shielding scheme is effective at reducing capacitive or electrically 
coupled noise. The same concerns regarding the difference in ground 
potentials, discussed in the Ground-Referenced Signal Connection section, 
also apply to this configuration.
For more information about the function of the NI PXI-4224 and other 
measurement considerations, refer to Chapter 4,