Calibration

Recommended warm-up time ................ 30 minutes

External calibration interval................... 1 year

Pre-Calibration Errors

1

Pre-calibration offset error
relative to input (RTI) ............................ 865 mV max

Signal conditioning
component only............................... ±50 mV typ, ±160 mV max

at a gain of 1

Pre-calibration gain error ....................... ±18,900 ppm max

Signal conditioning
component only............................... ±600 ppm typ, ±1,000 ppm max

at a gain of 1

Memory

FIFO buffer size ..................................... 512 samples

Data transfers ......................................... DMA, interrupts,

programmed I/O

DMA modes ........................................... Scatter-gather (single transfer,

demand transfer)

Configuration memory size.................... 512 words

Digital Triggers

Number of triggers ................................. 2

Purpose................................................... Start and stop trigger, gate, clock

Source..................................................... PFI 0/AI START TRIG

(front SMB connector),
PXI_TRIG<0..5> to PXI_Star
(PXI trigger bus)

Compatibility ......................................... 5 V/TTL

1

The pre-calibration errors apply only to users doing register level programming. Pre-calibration errors are not visible to