ARB Connector

The ARB connector provides the waveform output. The maximum
output levels on this connector depend on the type of load termination.
If the output of a DAQArb 5411 terminates into a 50

load, the output

levels are ±5 V, as shown in Figure 3-2. If the output of DAQArb 5411
terminates into a high impedance load (HiZ), the output levels are
±10 V. If the output terminates into any other load, the levels are:

where V

out

is the maximum output voltage level, R

L

is the load

impedance in ohms, and R

O

is the output impedance on the

DAQArb 5411. By default, R

O

= 50

, but the software can also set it

to 75

.

Note:

Software will set the voltage output levels based on a 50

load termination.

For more information on waveform generation and analog output
operation, refer to Chapter 4, Arb Operation. For specifications on the
waveform output signal, see Appendix A, Specifications.

Figure 3-2. Output Levels and Load Termination Using a 50

Output Impedance

V

out

=

±

x

10 V

R

L

R

L

+ R

O

ARB