Stepped trigger mode, Figure 411. continuous trigger mode for arb mode, Figure 412. continuous trigger mode for dds mode – National Instruments 5411 User Manual

Page 39: Stepped trigger mode -14

Advertising
background image

Chapter 4

Arb Operation

DAQArb 5411 User Manual

4-14

© National Instruments Corporation

You can use continuous trigger mode with the both the Arb and DDS
waveform generation modes, as follows:

Arb mode—Figure 4-11 uses the stages shown in Figure 4-4 to
illustrate a continuous trigger mode of operation for Arb waveform
generation mode.

Figure 4-11. Continuous Trigger Mode for Arb Mode

DDS mode—Figure 4-12 illustrates a continuous trigger mode of
operation for DDS waveform generation mode.

Figure 4-12. Continuous Trigger Mode for DDS Mode

Stepped Trigger Mode

After a start trigger is received, the waveform described by the first
stage is generated. Then, the device waits for the next trigger signal. On
the next trigger, the waveform described by the second stage is
generated, and so on. Once the staging list is exhausted, the waveform
generation returns to the first stage and continues in a cyclic fashion.

You can use the stepped trigger mode with the both the Arb and DDS
waveform generation modes, as follows:

Arb mode—Figure 4-13 uses the stages shown in Figure 4-4 to
illustrate a stepped trigger mode of operation for the Arb mode. If
a trigger is received while a stage is being generated, it will be
ignored. A trigger will be recognized only after the stage has been
completely generated.

Start Trigger

End of All Stages

End of All Stages

Repeat
Until Stopped

Start Trigger

End of All Stages

(f1,

T1

)

(f2,

T2

)

(f3,

T3

)

(f1,

T1

)

(f2,

T2

)

(f4,

T4

)

Repeat
Until Stopped