Figure 4-3. summary of analog input connections – National Instruments E Series User Manual

Page 49

Advertising
background image

Chapter 4 Signal Connections

National Instruments Corporation

4-13

DAQCard E Series User Manual

Figure 4-3 summarizes the recommended input configuration for both

types of signal sources.

Figure 4-3.

Summary of Analog Input Connections

Input

Differential

(DIFF)

Single-Ended —

Ground

Referenced

(RSE)

Single-Ended —

Nonreferenced

(NRSE)

Floating Signal Source

(Not Connected to Building Ground)

Grounded Signal Source

Examples
• Ungrounded Thermocouples

• Signal conditioning with isolated outputs
• Battery devices

Examples
• Plug-in instruments with
nonisolated outputs

+

-

+
-

V

1

ACH(+)

+

-

+
-

V

1

ACH(+)

ACH (-)

See text for information on bias resistors.

See text for information on bias resistors.

R

Signal Source Type

ACH (-)

AIGND

+

-

+
-

V

1

ACH

AIGND

+

-

+
-

NOT RECOMMENDED

ACH

+ V

g

-

V

1

Ground-loop losses, V

g

, are added to

measured signal

+

-

+
-

V

1

ACH

AISENSE

R

AIGND

+

-

+
-

V

1

ACH

AISENSE

AIGND

AIGND

Advertising