Figure 4-14. trig1 input signal timing, Figure 4-15. trig1 output signal timing – National Instruments E Series User Manual

Page 64

Advertising
background image

Chapter 4 Signal Connections

DAQCard E Series User Manual

4-28

National Instruments Corporation

Figures 4-14 and 4-15

show the input and output timing requirements

for the TRIG1 signal.

The DAQCard also uses the TRIG1 signal to initiate pretriggered data

acquisition operations. In most pretriggered applications, the TRIG1
signal is generated by a software trigger. Refer to the TRIG2 signal
description for a complete description of the use of TRIG1 and TRIG2
in a pretriggered data acquisition operation.

Figure 4-14.

TRIG1 Input Signal Timing

Figure 4-15.

TRIG1 Output Signal Timing

Rising-edge
polarity

Falling-edge
polarity

t

w

t

w

= 10 ns minimum

t

w

t

w

= 50-100 ns