Chapter 4 Signal Connections

DAQCard E Series User Manual

4-16

National Instruments Corporation

Differential Connections for Nonreferenced or
Floating Signal Sources

Figure 4-5 shows how to connect a floating signal source to a channel

on a DAQCard configured in DIFF input mode.

Figure 4-5 shows two bias resistors connected in parallel with the

signal leads of a floating signal source. If you do not use the resistors
and the source is truly floating, the source is not likely to remain within
the common-mode signal range of the PGIA, and the PGIA will
saturate, causing erroneous readings. You must reference the source to
AIGND. The easiest way is simply to connect the positive side of the
signal to the positive input of the PGIA and connect the negative side
of the signal to AIGND as well as to the negative input of the PGIA,

Figure 4-5.

Differential Input Connections for Nonreferenced Signals

+

-

+

Floating

Signal

Source

Input Multiplexers

Instrumentation

Amplifier

V

m

Measured

Voltage

-

V

S

-

+

I/O Connector

AIGND

Bias

Current

Return

Paths

Bias
Resistors
(see text)

ACH<8..15>

ACH<0..7>

AISENSE

Selected Channel in DIFF Configuration

PGIA