Chapter 4 Signal Connections

National Instruments Corporation

4-35

DAQCard E Series User Manual

mode, the first active edge disables the STARTSCAN signal, and the
second active edge enables STARTSCAN.

The AIGATE signal can neither stop a scan in progress nor continue a

previously gated-off scan; in other words, once a scan has started,
AIGATE does not gate off conversions until the beginning of the next
scan and, conversely, if conversions are being gated off, AIGATE does
not gate them back on until the beginning of the next scan.

SISOURCE Signal

Any PFI pin can externally input the SISOURCE signal, which is not

available as an output on the I/O connector. The onboard scan interval
counter uses the SISOURCE signal as a clock to time the generation of
the STARTSCAN signal. You must configure the PFI pin you select as
the source for the SISOURCE signal in the level-detection mode. You
can configure the polarity selection for the PFI pin for either active
high or active low.

The maximum allowed frequency is 20 MHz, with a minimum pulse

width of 23 ns high or low. There is no minimum frequency limitation.

Either the 20 MHz or 100 kHz internal timebase generates the

SISOURCE signal unless you select some external source. Figure 4-22
shows the timing requirements for the SISOURCE signal.

Figure 4-22.

SISOURCE Signal Timing

t

p

t

w

t

w

t

p

t

w

= 50 ns minimum

= 23 ns minimum