Single-module multiplexed scanning (indirect), Multiple-module multiplexed scanning, Figure 3-8 – National Instruments SCXI-1120 User Manual

Page 62: Single-module multiplexed scanning (direct), Figure 3-9

Advertising
background image

Chapter 3

Theory of Operation

© National Instruments Corporation

3-15

SCXI-1120 User Manual

except during changes from one Slot 0 scan list entry to the next, when SCANCON pulses high
to make the MUXCOUNTER reload its starting channel. Notice that although you are using
only a single module, you can put many entries with different counts in the Slot 0 FIFO, so that
some channels are read more often than others. You cannot change the start channel in the
module Configuration Register during a scan.

Cable

Assembly

Data Acquisition Board

Analog Input

Timing Output

TRIG0

SCANCON X

SCANCLK

MCH0

SLOT 0

Eight Isolated

Floating

Single-Ended

Inputs

SCXI-1000 or SCXI-1001 Chassis

SLOT X

SCXI-1120

Figure 3-8. Single-Module Multiplexed Scanning (Direct)

Single-Module Multiplexed Scanning (Indirect)

In this mode, the SCXI-1120 is not directly cabled to the data acquisition board. Instead, you
connect another module to the data acquisition board, and the analog output of the SCXI-1120 is
sent over Analog Bus 0, through the intermediate module, and then to the data acquisition board.
The SCXI-1120 receives its MUXCOUNTER clock from TRIG0, which is sent by the
intermediate module, as shown in Figure 3-9. Slot 0 operation is the same for direct connection
scanning.

Cable

Assembly

Data Acquisition Board

Analog Input

Timing Output

Other

Module

Analog Bus 0

SCANCON X

TRIG0

SCANCLK

MCH0

SCXI-1000 or SCXI-1001 Chassis

SLOT X

SLOT 0

SCXI-1120

Figure 3-9. Single-Module Multiplexed Scanning (Indirect)

Multiple-Module Multiplexed Scanning

In this mode, all the modules tie into Analog Bus 0 and SCANCON enables the output of their
amplifiers. The module that is directly cabled to the data acquisition board sends SCANCLK