Single-module multiplexed scanning (direct), Single-module multiplexed scanning (indirect), Channel scanning from other modules – National Instruments SCXI-1120 User Manual

Page 80: Multiple-module multiplexed scanning

Advertising
background image

Chapter 5

Programming

© National Instruments Corporation

5-9

SCXI-1120 User Manual

Single-Module Multiplexed Scanning (Direct)

To perform simple channel scanning, you must cable the SCXI-1120 to a data acquisition board.
See Chapter 2, Configuration and Installation, for more information.

To program the module for scanned-channel measurements, write the binary pattern

10XXXCCC

XX001101

to the SCXI-1120 Configuration Register.

CCC

represents the starting channel

number.

Single-Module Multiplexed Scanning (Indirect)

To indirectly scan a module, send the output of the scanned module onto Analog Bus 0, where it
is picked up by another module and transmitted to the data acquisition board.

Channel Scanning from Other Modules. To scan measurements from other modules, you must
cable the SCXI-1120 to a data acquisition board. See Chapter 2, Configuration and Installation,
for more information. The module programming steps are as follows:

1. Perform any necessary programming to ensure that no modules are driving Analog Bus 0.

For an SCXI-1120, clearing AB0EN in the Configuration Register will ensure that its output
is not driving AB0.

2. Write the binary pattern

10XXXXXX XX100011

to the SCXI-1120 Configuration Register.

This step disables the SCXI-1120 from driving Analog Bus 0 and allows Analog Bus 0 to
drive MCH0 through the output buffer.

3. Program the other module to be scanned.

Channel Scanning from the SCXI-1120 via Another Module. To scan the SCXI-1120 via other
modules, you must cable the other module to a data acquisition board, and the other module must
be able to transfer Analog Bus 0 to the data acquisition board. The other module must also be
able to send a SCANCLK*-compatible signal on TRIG0. See Chapter 2, Configuration and
Installation,
for more information. The module programming steps are as follows:

1. Perform any necessary programming to ensure that no modules are driving Analog Bus 0.

For an SCXI-1120, clearing AB0EN in the Configuration Register will ensure that its output
is not driving AB0.

2. Program the other module not to drive Analog Bus 0, but to send Analog Bus 0 to the data

acquisition board. Also program the other module to send a SCANCLK*-compatible signal
to TRIG0.

3. Write the binary pattern

01XXXCCC XX001111

to the SCXI-1120 Configuration Register,

where

CCC

is the starting channel number.

Multiple-Module Multiplexed Scanning

To scan multiple modules, you must connect one module to the data acquisition board, and the
module must be able to transfer Analog Bus 0 to the data acquisition board. This module must