Appendix a specifications, Analog input – National Instruments SCXI-1120 User Manual

Page 87

Advertising
background image

© National Instruments Corporation

A-1

SCXI-1120 User Manual

Appendix A
Specifications

This appendix lists the specifications for the SCXI-1120. These are typical at 25

°

C unless

otherwise stated. The operating temperature range is 0

°

to 50

°

C.

Analog Input

Gain (jumper-selectable)

1, 2, 5, 10, 20, 50, 100, 200, 500, 1,000, 2,000

Output range

±

5 V

Number of channels

8

Gain accuracy

0.15% of full scale

Offset voltage

Input

±

6

µ

V

Output

±

3 mV

Stability versus ambient temperature

Input offset drift

±

0.2

µ

V/

°

C

Output offset drift

±

200

µ

V/

°

C

Gain drift

20 ppm/

°

C

Input bias current

±

80 pA

Input resistance

Normal

40 M

Power off

50 k

Overload

50 k

Output resistance

Multiplexed-Output mode

100

Parallel-Output mode

330

Filtering (jumper selectable)

4 Hz (-10 dB) or 10 kHz (-3 dB), 3-pole RC

Output selection time

(with 5 V step, all gains)
0.012% accuracy

1

5.2

µ

sec typical, 7

µ

sec maximum

0.006% accuracy

2

10

µ

sec

0.0015% accuracy

2

20

µ

sec

1

Includes the combined effects of the SCXI-1120 and the AT-MIO-16F-5

2

Includes the combined effects of the SCXI-1120 and the AT-MIO-16X

Advertising