Register Descriptions

Chapter 4

SCXI-1120 User Manual

4-8

© National Instruments Corporation

FIFO Register

The FIFO Register is used to add entries to the Slot 0 FIFO. The FIFO contains the Slot 0 scan
list. Each entry contains a slot number to be accessed, and a count number to determine the
number of samples to be taken from that slot. To write to the FIFO Register, follow the
procedure given in the Register Writes section in Chapter 5, Programming, using 14 as the slot
number, and writing 16 bits to the FIFO Register. The register will shift in the data present on
the MOSI line, bit 7 first, when Slot 14 is selected by the Slot-Select Register. The Slot 0 scan
list is created by consecutive writes to the FIFO Register. Each write creates a new entry at the
end of the scan list. The maximum number of entries is 256. To clear the FIFO of all entries,
clear the HSRS* bit in the HSCR.

Type:

Write-only

Word Size:

16-bit

Bit Map:

15

14

13

12

11

10

9

8

X

X

X

X

X

MOD3

MOD2

MOD1

7

6

5

4

3

2

1

0

MOD0

CNT6

CNT5

CNT4

CNT3

CNT2

CNT1

CNT0

Bit

Name

Description

15-11

X

Don't care bits – Unused.

10-7

MOD<3..0>

Module Number – The value of these bits plus one
determines the number of the slot to be accessed for
this scan entry. For example, to access Slot 6,
MOD<3..0> would be 0101.

6-0

CNT<6..0>

Count – The value of these bits plus one determines
how many samples will be taken before the next
scan list entry becomes active. A value of zero
corresponds to one sample and a value of 127
corresponds to 128 samples.