Shock and vibration, Safety, Electromagnetic compatibility – National Instruments NI VXIpc-882 User Manual
Page 84

Appendix A
Specifications
© National Instruments Corporation
A-3
Shock and Vibration
Operational shock
1
................................. 30 g peak, half-sine, 11 ms pulse
(Tested in accordance with
IEC-60068-2-27. Test profile
developed in accordance with
MIL-PRF-28800F.)
Random vibration
Operating
2
....................................... 5 to 500 Hz, 0.3 g
rms
(with solid-state hard drive)
Nonoperating .................................. 5 to 500 Hz, 2.4 g
rms
(Tested in accordance with IEC-60068-2-64. Nonoperating test profile
exceeds the requirements of MIL-PRF-28800F, Class B.)
Note
Specifications are subject to change without notice.
Safety
This product meets the requirements of the following standards of safety
for electrical equipment for measurement, control, and laboratory use:
•
IEC 61010-1, EN 61010-1
•
UL 61010-1, CSA 61010-1
Note
For UL and other safety certifications, refer to the product label or the
section.
Electromagnetic Compatibility
This product meets the requirements of the following EMC standards for
electrical equipment for measurement, control, and laboratory use:
•
EN 61326 (IEC 61326): Class A emissions; Basic immunity
•
EN 55011 (CISPR 11): Group 1, Class A emissions
•
AS/NZS CISPR 11: Group 1, Class A emissions
•
FCC 47 CFR Part 15B: Class A emissions
•
ICES-001: Class A emissions
1
Only valid with an NI VXIpc-882 using a solid-state drive, not with the shipping configuration, which contains a hard disk
drive.
2
Only valid with an NI VXIpc-882 using a solid-state drive, not with the shipping configuration, which contains a hard disk
drive.