2.4 Location, Mounting, and Ventilation

Use the power supply in rack mounted or in benchtop applications.

2.4.1 Rack

Mounting

WARNING

Ensure that the screws for the rack mount brackets do not extend more than
1/8" (3.0 mm) into the sides of the unit.

The power supply is designed to fit in a standard 19" (483 mm) equipment rack. To install:

1.

Use the rack mount brackets at either side of the front panel to install the power supply in a rack. To
mount the supply for relay rack applications, remove the brackets from the location near the front panel
and re-install them at their alternate locations which are midway along each side.

2.

Provide adequate support for the rear of the unit without obstructing the ventilation inlets on the sides
of the unit. Use a support bar such as Hammond RASB19WH2 at the rear of the unit. Follow the
manufacturer's instructions to install the support bar.

To install the 2800 Watt power supply with rack mount slides, use a typical slide kit such as ZERO C-300 S-18.
Follow the manufacturer's instructions to install the slides. Add a 1/8" (3.0 mm) spacer between the chassis and
the slide.

2.4.2 Ventilation

Whether you place the power supply in a rack or on a bench, allow cooling air to reach the ventilation inlets on
the sides of the unit and allow 4" (10 cm) of unrestricted air space at the rear of the unit for the fan exhaust. Any
ventilation space at the top and bottom of the supply will further lower internal operating temperatures.

Check Section 1.3.6 for the operating ambient temperature range and the operating altitude specification.