Valor™ 7000, Series en-13, 2 cal menu – Ohaus Valor 7000 Series COMPACT FOOD SCALES Manual User Manual

Page 15

Advertising
background image

Valor™ 7000

Series EN-13


4.2 Cal Menu

Enter this menu to perform calibrations.

Initial Calibration

When the scale is operated for the first time, a span calibration is recommended to ensure

accurate weighing results. Before performing the calibration, be sure to have the appropriate

calibration weights as listed in table 4-1. Ensure that the LFT switch / calibration lock is set to

unlock position. See figure 5-1.
Or adjust the GEO setting according to your location (see table 4-2).
Procedure:
Press and hold Menu until [mMeNU] (Menu) is displayed. When the button is released, the display

will show [C.A.L]. Press Yes to accept. [SpaN] will then be shown. Press Yes to begin the span

calibration. [ 0

kg

] will be displayed. Press Yes to accept. [--C--] will be displayed while zero

reading is stored. Next, the display shows the calibration weight value. Place the specified

calibration mass on the pan. Press Yes to accept the weight or No to select an alternate weight.

[--C--] will be displayed while the reading is stored. The display will show [done] if the calibration

was successful. The scale returns to the previous application mode and is ready for use.

TABLE 4-1

Required Span Calibration Mass (sold separately)

Capacity Mass

*

Capacity Mass

*

1500g 1.5kg / 3lb

15000g 15kg / 30lb

3000g 3kg / 6lb

30000g 30kg / 60lb

6000g 6kg / 15lb

Notes: When active unit is g or kg, the calibrating unit will be in kg.

When active unit is lb, oz or lb:oz, the calibrating unit will be in lb.

For linearity calibration, the calibration Mass is fixed.

Mid-point is always half of full capacity.








Span

[SpaN]

Initiates a span calibration procedure (zero and span)

Lin

[LIN]

Initiates a linearity calibration procedure (zero, mid-point and span).


GEO

[GEO]

Geographical Adjustment Factor (GEO) is used to adjust the calibration based on the current

location. Settings from 0 to 31 are available with 12 being the default.

Refer to table 4-2 to determine the GEO factor that corresponds to your location.

End Cal

[End]

Advance to the next menu or return to the top of the current menu.

Span:

Perform

Linearity:

Perform

Geographic:

Adjustment:

Set 0...12...31*

End Calibration: Exit menu

*Bold always represents factory default value.