Figure 5-1. Sealing

5.3.2 Audit Trail Seal (USA and Canada only)

For jurisdictions that use the audit trail sealing method, the local weights and measures official

or authorized service agent must record the current configuration and calibration event counter

values at the time of sealing. These values will be compared to values found during a future

inspection.
Note:

A change to an event counter value is equivalent to breaking a physical seal.

The audit trail uses two event counters to record changes to configuration and calibration

settings.

The configuration event counter (CFG) will index by 1 when exiting the menu if one or

more of the following menu settings are changed: Zero Range, Stable Range, Auto Zero

Tracking (AZT), Units (kg, g, oz, lb or lb:oz) or Stable Only.

The calibration event counter (CAL) will index by 1 when exiting the menu if a Span

Calibration or GEO setting change is made. Note that the counter only indexes once,

even if several settings are changed.

The event counters can be viewed by pressing and holding the MENU button.

While the button is held, the display will show MENU followed by Audit.

Release the button when Audit is displayed to view the audit trail information.

The audit trail information is displayed in the format CFGxxx and CALxxx.

The scale then returns to normal operation

Paper Seal

Wire Seal

Lock Switch

Unlocked Locked with Paper Seal Locked with Wire Seal