5. LEGAL FOR TRADE

When the scale is used in trade or a legally controlled application it must be set up, verified and

sealed in accordance with local weights and measures regulations. It is the responsibility of the

purchaser to ensure that all pertinent legal requirements are met.

5.1 Capacity Label

A label showing the capacity and readability of the scale must be installed near each display. If

the Capacity Labels were installed prior to delivery, no further action is needed. If the Capacity

Labels were not installed, they have been placed in the packaging material. Affix the labels

above the displays as shown in Figure 5-1.
Note:

The Capacity Labels will be destroyed upon removal, so only attempt to install them once.

Figure 5-1. Front and rear display Capacity Label locations

5.2 Settings

Before verification and sealing, perform the following steps:

1. Verify that the menu settings meet the local weights and measures regulations.

2. Perform a calibration as explained in Section 4.2.

3. Set the switch to Locked. See figure 5-2.

The Menu Lock switch limits changes to the Cal, Setup, Readout, Unit and Print menus. The

switch in type approved models may set some scale settings as required by the approval

agency. The switch may be secured using paper or wire seals.


Note:

When LEGAL FOR TRADE is set to ON (LFT Switch in locked position), the menu

settings are affected as follows:

Calibration (C.A.L) menu is not accessible

Zero Range is locked at 2%

Stable Range setting is locked at 1d

Auto-Zero Tracking setting is locked at 0.5d

IR Function, Filter and Units are locked at their current settings

Stable Only is locked On

Auto Print/ Continuous is disabled

Lb:oz is locked at Off.


5.3 Verification and Sealing

The local weights and measures official or authorized service agent must perform the

verification procedure.

5.3.1 Physical Seals

For jurisdictions that use the physical sealing method, the local weights and measures official or

authorized service agent must apply a security seal to prevent tampering with the settings.

Refer to the illustration below for sealing methods.

Capacity Label