3.3.7 Calibration Test

Calibration test feature allows a check of a known calibration mass against the last stored calibration information
in the balance. Sample display illustrates an 8.1 kg balance.

Procedure

• Press the Setup button, CAL is displayed.

• Press Enter button, CAL TYPE is displayed.

• Press Enter button, CAL TYPE InCAL is displayed.

• Press

button to select CALTEST calibration,

CALtESt is displayed.

• Press Enter button, WORKING is displayed.

NOTE: If a weight is left on the pan, the balance will
display CLR PAN (remove the weight from the pan). The
balance automatically resumes calibration.

• Display changes to PUT WT 8000 g. The displayed

weight is the full capacity of the balance.

• Place specified calibration mass on pan.

• Press Enter button, WORKING is displayed. After a

few seconds, DIFF is displayed. The display now
indicates the actual difference in weight between
what value was just placed on the pan and the
previous weight value which was stored in the
balance. After approximately 8 seconds, the display
returns to the WEIGH mode.

• Remove calibration test mass from the pan.

Setup

Use keys to change,

press Enter

Change

Use keys to change,

press Enter

Change

Setup

Setup

Use keys to change,

press Enter

Change

Setup

press Enter

Change

g

Setup

g

Change

Mode

Units

Setup

Print

Enter

Setup

Use keys to change,

press Enter

Change

Mode

Units

Setup

Print

Enter

Stable

g

Advertising