Legal for trade (lft), 7 legal for trade (lft) – Ohaus E1RW60 Explorer Balances User Manual

Page 82

Advertising
background image

28

Procedure

• Press the Setup button, CAL is displayed.

• Press

button repeately until LFT is displayed.

• Press Enter button, LFT NET is displayed.
• Press Enter button,LFT NET OFF is displayed.

• Press

or

button and select either ON or

OFF.

• Press Enter button, SAVED is momentarily dis-

played, then LFT LOCK is displayed.

• Press Enter button, LFT LOCK OFF is displayed.

• Press

or

button and select either ON or

OFF.

• Press Enter button, SAVED is momentarily displayed

followed by EXIT.

• Press Enter button, balance returns to weigh mode.

NOTE: For legal for trade applications, the balance must
be physically sealed. Refer to section on LFT Sealing.

When the balance is first turned ON and LFT has been
previously set ON, the following display will appear if
LFT is set in the menu and the Lock Switch is set ON.

Mode

Units

Setup

Print

Enter

4.7 Legal for Trade (LFT)

Legal for Trade (LFT) is a software controlled option which can be set to LFT NET and LFT LOCK. When set to
LFT NET, the balance will indicate a NET weight during legal for trade operation. When LFT LOCK is set ON,
certain items in the Calibration, Readout, Print, Mode and Units menus are automatically preset and locked to
permit the balance to operate in a legal for trade application and works in conjunction with a Lock Switch. Default
setting is UNLOCKED. See default table.

LFT and Lock Switch

Menu lock

Default Value

Date Menu

Unlocked

Time Menu

Unlocked

Calibration Menu
Span, Linearity, User

Locked

CalTest

Unlocked

Internal Calibration

Unlocked

Cal Adj

Locked

Cal Msg

Locked ON

Readout Menu
Stability

Unlocked

1d (limited to .5d and 1d)

Auto zero

Unlocked

1d (limited to OFF and .5d)

Filter Level

Unlocked

-2-

GLP Data Menu

Unlocked

GLP Selections

Unlocked

Print Options

Unlocked

RS232 Menu

Unlocked

LFT Menu

Lockswitch Locked

Function menu

Locked

Weigh

Units Menu

Locked

Grams*

Global Menu

Locked

Custom Menu

Unlocked

* Units oz and oz t cannot be simultaneously enabled.

If Print Numeric Data is turned ON, then Print Stable Data Only is
locked ON.

DEFAULT TABLE

Setup

Use keys to change,

press Enter

Change

The display check countdown appears during the first
120 seconds after plugging in the balance and only
when the balance has been previously set with Type
Approved/Legal for Trade ON.

When the balance is first turned ON and LFT has been
previously set ON, the following display will appear if
LFT is set in the menu and Calibration menu is locked,
and the Lock Switch is set ON.

Stable

NET

Center
of Zero

g

When LFT NET and LFT LOCK are both turned ON, the
weigh mode display appears as shown below.

Advertising