Rs232, Rs232 interface, 2 rs232 interface – Ohaus E1RW60 Explorer Balances User Manual

Page 90: Rs232 command table

Advertising
background image

36

RS232 COMMAND TABLE

Command

Command

Command

Command

Command
Character

Character

Character

Character

Character

Description

Description

Description

Description

Description

?

?

?

?

?

Print current mode

nnnnA

nnnnA

nnnnA

nnnnA

nnnnA

Set Auto Print feature to “nnnn”
(see table).

Field: Mode Stab CR

LF

Length:

5

1

1

1

blank if stable
“ ? ” if unstable

nnn = 0

Turns feature OFF

nnn = S

Output on stability

nnn = C

Output is continuous

nnnn = 1-3600 Sets Auto Print

Interval

5.2 RS232 Interface

Explorer balances are equipped with a bidirectional
RS232 compatible interface for communication with
printers and computers. When the balance is connected
directly to a printer, displayed data can be output at any
time by simply pressing PRINT, or by using the Auto
Print feature.

Connecting the balance to a computer enables you to
operate the balance from the computer, as well as
receive data such as displayed weight, weighing mode,
stability status, etc.

The following sections describe the hardware and
software provided with the balance

Hardware
On the rear of the balance, the right-hand, 9-pin male
subminiature “D” connector is provided for interfacing
to other devices. The pinout and pin connections are
shown in the adjacent illustration.

The balance will not output any data unless pin 5 (CTS)
is held in an ON state (+3 to +15 V dc). Interfaces not
utilizing the CTS handshake may tie pin 5 to pin 6 to
defeat it.

Output Formats
Data output can be initiated in one of three ways: 1) By
pressing PRINT; 2) Using the Auto Print feature; 3)
Sending a print command (“P”) from a computer.

The output format is illustrated in the RS232 command
table which follows.

RS232 Connection

1

N/C

2

Data Out (TXD)

3

Data In (RXD)

4*

Tare (External signal)

5

Clear To Send (CTS)

6

Data Terminal Ready (DTR)

7

Ground

8

Request To Send (RTS)

9*

Print (External signal)

* External PRINT and/or TARE

switches may be installed
as shown in the diagram.
Momentary contact
switches must be used.

1

2

3

4

5

6

7

8

9

PRINT *

TARE *

RS232 Commands
All communication is accomplished using standard ASCII
format. Characters shown in the following table are
acknowledged by the balance. Invalid command re-
sponse "ES" error indicates the balance has not recog-
nized the command. Commands sent to the balance
must be terminated with a carriage return (CR) or car-
riage return-line line feed (CRLF). For example, a tare
command should appear as shown in the adjacent dia-
gram. Data output by the balance is always terminated
with a carriage return - line feed (CRLF).

Rear of Balance

mg
g
kg
dwt
ct
oz
oz t

GN
tael
tael
tael
momme

lb

N
tical
custm
Pcs
%

Advertising